86
Respiration
IM Respiration
Technique Impedance Pneumography
Range 0, 3 to 120 breaths/min
Accuracy ±3 breaths/min
Leads RA to LA
Lead Off Condition Detected and displayed
Defibrillator Protection Protected
AW Respiration
Technique Nondispresive Infrared Spectroscopy
Range 0 to 50 breaths/min
Accuracy ±1 breaths/min
Display Sweep Speeds 6.25 mm/s, 12.5 mm/s, 25.0 mm/s
Wave Size Auto, Level 1~8
NIBP
Pulse Rate
Pulse Rate Range Adult/Pediatric 40 to 200 BPM
Neonatal
40 to 240 BPM
Pulse Rate Accuracy ±2 BPM or ±2%, whichever is greater
NIBP (Non-Invasive Blood Pressure)
Technique Oscillometric Measurement
Measurement Modes MANUAL, AUTO and CONT
NIBP AUTO Mode
Intervals
Off, 1, 2.5, 3, 5, 10, 15, 30, 60, 90 minutes
Measurement Range Adult/Pediatric
SYS
60 to 250 mmHg
MAP 45 to 235 mmHg
DIA 40 to 200 mmHg
Neonatal
SYS
40 to 120 mmHg
MAP 30 to 100 mmHg
DIA 20 to 90 mmHg
NIBP Accuracy Mean error and standard deviation per ANSI/AAMI
SP10:2002+A1:2003
Pressure Display Range Adult/Pediatric 0 to 300 mmHg
Neonatal
0 to 150 mmHg
Pressure Display
Accuracy
Within ±3mmHg
Initial Cuff Inflation Adult/Pediatric
120, 140, 160, 180, 200, 220, 240, 260, 280 mmHg
(16.0, 18.7, 21.3, 24.0, 26.7, 29.3, 32.0, 34.7, 37.3 kPa)
Neonatal
80, 90, 100, 110, 120, 130, 140 mmHg
(10.7, 12.0, 13.3, 14.7, 16.0, 17.3, 18.7 kPa)
Automatic Cuff Deflation Measurement time exceeding 180s in adult/pediatric
(90s in neonatal) or maximum pressure value exceeding
300 mmHg in adult (150 mmHg in neonatal).