82
Electrical
Instrument
Power Requirements AC Power
100 to 240VAC, 50/60 Hz, 110VA
Fuses qty 2, T3.15 A, 250 volts / T5.0 A, 250 volts
Battery
Type Li-ion battery
Operating time 1 hour (2200mAh), 3 hours (4400mAh), 5 hours
(6600mAh)
At the following condition:
no printing
no external communication
no audible alarm sound
one NIBP measurement per 15 minutes at 25°C
Voltage/Capacity 11.1 V / 2200 mAh (1 hour type)
11.1 V / 4400 mAh (3 hours type)
11.1 V / 6600 mAh (5 hours type)
Recharge Over 4, 8 or 12 hours with monitor turned on/off
Life Cycle 6 months, new battery fully-charged
After 2 months storage the M30 would run for 50% of
stated battery life.
Compliance 91/157/EEC
Environmental Conditions
Operation
Temperature 5 to 40°C (41 to 104°F)
Humidity 30 to 85% RH, non-condensing
Altitude 700 to 1060 hPa
Transport and Storage (in shipping container)
Temperature 20°C to 60°C (4°F to 140°F)
Humidity 10 to 95% RH, non-condensing
Altitude 500 to 1060hPa
Note: The system may not meet its performance specifications if stored or used
outside the specified temperature and humidity range.