Measurement Parameters
ECG
Heart Rate
Measurement Range 0, 30 to 300 BPM
Accuracy ±1 BPM or ±1% whichever is greater
Average Response Time 5 seconds (from 80 to120 BPM)
9 seconds (from 80 to 40 BPM)
Tall T-wave Rejection maximum T-wave amplitude 1.8 mV
ECG (Electrocardiograph)
Leads 3 / 5 Lead
Lead I, II, III, aV
R, aVL, aVF, V (Chest Lead)
Lead Off Detection Detected and displayed
Pacer Detection Detect pacer pulses of ±2mV to ±700mV with pulse
widths of 0.1 to 2msec and rise times 10% of width
not to exceed 100msec
Input
Input Impedance 5 M ohm or more
Input Dynamic Range ±5 mV AC, ±300 mV DC
Voltage Range ±0.5 mV ~ ±5 mV
Signal Width 40 to 120 ms (Q to S)
Output
Frequency Response (Bandwidth)
Low Extend 0.05 to 40 Hz
Filter 0.5 to 30 Hz
Monitor 0.5 to 40 Hz
Hum filter 50 Hz and 60 Hz
ECG Size Auto, 1.25, 1.7, 2.5, 5.0, 7.5, 10.0, 15.0, 20.0 mm/mV
Display Sweep Speeds 12.5 mm/sec, 25.0 mm/sec, and 50.0 mm/sec
Display Sensitivity 10 mm/mV (×1)
Pacing Pulse Detection On, Off
ST Level Analysis On, Off
Arrhythmia Detection On, Off
Electrode Disconnect Alarm Display and/or sound
CMRR 80 dB or more
Defibrillator Discharge
Recovery
<5 sec per IEC60601-2-27
Defibrillator Protection Protected
ECG (Arrhythmia Supplemental Information as required by AAMI EC13)
Heart rate calculation Heart rate is computed by averaging the 20 most
recent RR intervals if there is more than 21 QRS for
10 sec.
If there is less than 21 QRS, heart rate is computed
by averaging RR intervals occurring for 10 sec.