TROUBLESHOOTING
General
This section provides information that can be helpful in troubleshooting the M30 monitor.
How to Use This Section
If the Unit is not functioning properly, please check on the following items before calling
for repair service. Use this section in conjunction with the Performance Verification
section and the Spare Parts section. To remove and replace a part suspected to be
trouble, follow the instructions in the Disassembly Guide section.
Who Should Perform Repairs
Only qualified service personnel should open the monitor housing, remove and replace
components, or make adjustments in accordance with this service manual. If your medical
facility does not have qualified service personnel, contact Mediana Technical Service
Department.
Replacement Level Supported
The replacement level supported for this product is to the printed circuit board (PCB
assembly) and major subassembly level. Once you isolate a suspected PCB assembly,
follow the procedures in the Disassembly Guide section to replace the PCB assembly
with a known good PCB assembly. Check to see if the trouble symptom disappears and
that the monitor passes all performance tests.
If the trouble symptom persists, swap back the replacement PCB assembly with the
suspected malfunctioning PCB assembly (the original PCB assembly that was installed
when you started troubleshooting) and continue troubleshooting as directed in this section.
Obtaining Replacement Parts
Mediana provides technical assistance information and replacement parts. To obtain
replacement parts, contact Mediana Technical Service Department. Refer to the part
names and part numbers listed in the Spare Parts section.