Air Leakage Test
Figure 6. Air Leakage Test
1. Ensure the monitor is set up with dummy can-large.
2. Ensure NIBP Test Mode screen is active on the monitor, then select Air Leakage Test
by the jog dial.
3. The monitor displays the pressure of approximately 290 mmHg automatically.
4. The test result displays at the test completion. The initial pressure value at 1 minute is
displayed after the test start and the air leakage value at further 3 minutes after the 1
minute elapsed.
Note: The test will have been successfully completed if the pressure has dropped by 6
mmHg, or less, during the 1-minute period.
Inflation Time Measurement
Figure 7. Inflation Time Measurement