
15
pulse rate tone decreases.
10. Set QRS volume to Off and return to the monitoring screen. Verify beeping pulse rate
tone is no longer audible.
11. Return QRS volume to a comfortable level.
Sensor LED Test
This procedure uses normal system components to test circuit operation. An SpO
2
sensor,
DS-100A is used to examine LED intensity control. The red LED is used to verify intensity
modulation caused by the LED intensity control circuit.
1. Connect the monitor to an AC power source.
2. Press the Power Button to turn on the monitor.
3. Connect the SpO
2
extension cable to the monitor.
4. Connect the SpO
2
sensor to the SpO
2
extension cable.
5. Leave the sensor open with the LEDs and photodetector visible.
6. After the monitor completes its normal power-up sequence, verify that the sensor LED
is brightly lit.
7. Slowly move the sensor LED in proximity of the photodetector element of the sensor
(close the sensor slowly). Verify as the LED approaches the optical sensor, the LED
intensity decreases.
8. Open the sensor and notice that the LED intensity increases.
9. Repeat step 7 and intensity will again decrease. This variation is an indication that the
microprocessor is in proper control of LED intensity.
10. Press the Power Button to turn off the monitor.
Restoring Power-On Default Settings
The following test procedures will verify that alarms are activated at the level of factory
default alarm limits and that any changed settings are saved and in effect when the user
changes alarm limit settings and saves the current settings as a power default.
Table 2. Parameter Alarm Limit Factory Defaults
Alarm Conditions Adult Pediatric Neonatal
HR/PR Upper Alarm Limits 120 bpm (beats per minute) 160 bpm 200 bpm
HR/PR Lower Alarm Limits 50 bpm 75 bpm 100 bpm
ST Level (1 to 5) High Alarm Limits 0.50mV 0.50mV 0.50mV
ST Level (1 to 5) Low Alarm Limits -0.50mV -0.50mV -0.50mV
Arrhythmia High Alarm Limits 5 /min 5 /min 5 /min
NIBP SYS Upper Alarm Limits 160mmHg 120 mmHg 90 mmHg
NIBP SYS Lower Alarm Limits 90 mmHg 70 mmHg 40 mmHg
NIBP DIA Upper Alarm Limits 90 mmHg 70 mmHg 60 mmHg
NIBP DIA Lower Alarm Limits 50 mmHg 40 mmHg 20 mmHg
NIBP MAP Upper Alarm Limits 110 mmHg 90 mmHg 70 mmHg
NIBP MAP Lower Alarm Limits 60 mmHg 50 mmHg 30 mmHg
%SpO
2
Upper Alarm Limits 100 % 100 % 100 %
%SpO
2
Lower Alarm Limits 90 % 90 % 85 %
RESP Upper Alarm Limits 30 bpm (breaths per minute) 30 bpm 100 bpm
RESP Lower Alarm Limits 8 bpm 8 bpm 30 bpm
Comentários a estes Manuais