35
SERVICE MENU AND FACTORY DEFAULT
General
This section discusses use of the Service menu to configure ‘Save Current Values as
Power-On Default’, ‘Save Factory Default Setting as Power-On Default’, ‘Alarm
Suspension Period’, ‘Alarm Silence Period’, ‘Alarm Reminder Tone’, ‘Audible Alarm Type’,
‘Language’, ‘NIBP TEST MODE’ and ‘System Information'. Also, this section explains
briefly the factory default settings.
Service Menu
The purpose of the Service menu is to allow the authorized user to create a Power-On
Default for the settings in effect each time the monitor is powered on. Once the Service
Menu is entered, physiological monitoring is terminated. The screen layouts do not display
any information associated with normal monitoring operation. Use the following procedure
to configure the Service Menu for the monitor (also see the Using the Monitor section of
the operation manual):
Figure 9. The access of Service Menu via Setup Menu
1. Set the monitor to normal monitoring mode.
2. Rotate the jog dial to highlight the Setup icon located on the bottom of the screen and
then press the jog dial. Setup Menu displays.
3. Rotate the jog dial to highlight Service Menu in Setup Menu, and then press the jog
dial to access the Service Menu.
4. Three digits are displayed in the Level 2 Menu as shown in Figure 10.