SAFETY TEST
TEST CONDITIONS LIMIT (uA) RESULTS REMARKS
Patient auxiliary current ECG LL-SpO
2
(OS) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LL-SpO
2
(SFC OE) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LL-SpO
2
(NCRM) 10 Pass / Fail Value: uA
Patient auxiliary current ECG LL-SpO
2
(SFC OSRM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LL-SpO
2
(SFC OERM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG RA-SpO
2
(NC) 10 Pass / Fail Value: uA
Patient auxiliary current ECG RA-SpO
2
(OS) 50 Pass / Fail Value: uA
Patient auxiliary current ECG RA-SpO
2
(SFC OE) 50 Pass / Fail Value: uA
Patient auxiliary current ECG RA-SpO
2
(NCRM) 10 Pass / Fail Value: uA
Patient auxiliary current ECG RA-SpO
2
(SFC OSRM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG RA-SpO
2
(SFC OERM) 50 Pass / Fail Value: uA
Patient auxiliary current TEMP-SpO
2
(NC) 10 Pass / Fail Value: uA
Patient auxiliary current TEMP-SpO
2
(OS) 50 Pass / Fail Value: uA
Patient auxiliary current TEMP-SpO
2
(SFC OE) 50 Pass / Fail Value: uA
Patient auxiliary current TEMP-SpO
2
(NCRM) 10 Pass / Fail Value: uA
Patient auxiliary current TEMP-SpO
2
(SFC OSRM) 50 Pass / Fail Value: uA
Patient auxiliary current TEMP-SpO
2
(SFC OERM) 50 Pass / Fail Value: uA
Remarks
NC: Normal Condition
NCRM: Normal Condition Reverse
SFC: Single Fault Condition
OS: Single Fault Condition (Open Line/Neutral)
OSRM: Single Fault Condition (Open Line/Neutral) Reverse
OE: Single Fault Condition (Open Earth)
OERM: Single Fault Condition (Open Earth) Reverse