Alarm Conditions Adult Pediatric Neonatal
TEMP Upper Alarm Limits 39.0°C (102.2°F) 39.0°C
(102.2°F)
39.0°C
(102.2°F)
TEMP Lower Alarm Limits 36.0 °F (96.8 °F) 36.0 °F
(96.8 °F)
36.0 °F
(96.8 °F)
EtCO
2
Upper Alarm Limits 80 mmHg 80mmHg 80mmHg
EtCO
2
Lower Alarm Limits 0 mmHg 0mmHg 0 mmHg
InCO
2
Upper Alarm Limits 20 mmHg 20 mmHg 20 mmHg
InCO
2
Upper Alarm Limits 0 mmHg 0 mmHg 0 mmHg
1. Turn on the monitor under Factory default settings.
2. Select the Alarm Limits Icon to display the Alarm Limits Menu.
3. Verify that alarm limits are set as shown in Table 2.
4. Change Patient mode in the setup menu from Adult to Pediatric or Neonatal, then
verify that alarm limits are set as shown in Table 2.
5. Change alarm limit value via Alarm Limits Menu.
6. Save the current changed alarm limit values as a power on default setting via the
service menu (see Service Menu and Factory Defaults section).
7. Select “Done” to turn off the monitor.
8. Press Power Button to turn on the monitor.
9. Verify alarm limits are set to the current changed alarm limit values.
Printer Test (Option)
If an optional printer is installed in the monitor, the following test procedures will verify the
printer performance.
1. Connect the monitor to an AC power source.
2. Press the Power Button to turn on the monitor.
3. Connect all necessary simulators to the monitor.
4. Select the Setup Icon to display the Setup Menu.
5. Test #1: One-Shot printing
a. Set Print Mode to One-Shot via the Setup Menu.
b. Press the Print Button when all the parameter signals display normally.
c. Verify that the parameter values and waveforms are printed out for 20
seconds.
6. Test #2: Continuous printing
a. Set Print Mode to Continuous via the Setup Menu.
b. Press the Print Button when all the parameter signals display normally.
c. Verify the parameter values and waveforms are printed out continuously.
d. Verify printing stops when the Print Button is pressed again.
7. Test #3: Print Speed
a. Set Print Speed to 25 mm/s.