Pressure Sensor Accuracy Test
Figure 5. Pressure Sensor Accuracy Test
1. Ensure Bio-Tek simulator is in the static pressure test mode.
2. The NIBP test screen is active on the monitor, then select Pressure Sensor Accuracy
Test by the jog dial.
3. Press Select button on the simulator until simulator displays “Pressure Source Set Test
Pressure”. Adjust the pressure on the simulator for 250, 150, 50 and 0 mmHg.
4. Press Start Pump button on the simulator. The simulator will begin to pressurize. Allow
15-20 seconds for pressure to stabilize.
5. The current pressure in mmHg will be displayed on both of the simulator and the
monitor displays. Ensure the monitor pressure sensor accuracy meets the performance
standard of ANSI/AAMI SP-10:2002+A1:2003 (within the specification by more than ±3
mmHg or 2 percent of reading, whichever is greater) to successfully complete the test.