43
FIRMWARE DOWNLOAD
General
This section is for the purpose of reloading Firmware into the monitor when the
possibility of corrupted Firmware exists, or updating Firmware with a new system
revision (system/device version). Call Mediana Technical Service Department for the
latest version of Firmware utility required.
Note: The firmware download can be performed via USB or TCP/IP. If the USB and
TCP/IP are connected to the monitor simultaneously, the firmware downloading
would be performed by the method detected first.
Equipment Needed
Table 13 lists the equipment required for Firmware download.
Table 13. Required Equipment for Firmware Download
Equipment Description
USB
Firmware Download Tool USB Memory with M30 Field Utility
(firmware downloading software)
TCP/IP
Firmware Downloading Cable LAN Cable
Firmware Downloading Software M30 Field Utility
Personal Computer (PC) With TCP/IP Port
How to Download
1. Turn off the monitor.
2. Turn on the monitor by pressing the Home Button and NIBP Start/Stop Button
simultaneously.
3. The monitor will display the firmware upgrade screen as shown in Figure 12.
Figure 12. Firmware Downloading Display