38
Save Current Values as Power-On Default
The current settings become the power-up defaults next time the unit is powered up.
Save Factory Default Setting as Power-On Default
The factory default settings become the power-up defaults next time the unit us powered
up. See Table 12.
Alarm Suspension Period
If the Alarm Suspension Period is set to anything other than OFF or Indefinite, the
audible alarm is not activated for the specified time interval by pressing and holding the
Alarm Stop Button for 2 seconds. If OFF is selected, the audible alarm suspension is not
allowed to activate. If Indefinite is selected, the audible alarm suspension continues until
canceled.
Audible Alarm Silence Period
Pressing the Alarm Stop Button temporarily silences alarms for the period selected in
the Service Menu. The factory default alarm silence period is 60 seconds.
Alarm Reminder Tone
The Alarm Reminder Tone menu is activated with the user-selectable period in the
Service Menu. The interval can be set to OFF, 3 or 10 minutes. If OFF is selected, the
reminder tone will be disabled.
Audible Alarm Type
GN924 or IEC60601-1-8
The M30 has two different audible alarm types, called GN924 and IEC60601-1-8. They
have different tone pitch and on-off beep patterns. (Refer to Alarms and Limits section in
the operator’s manual)
Language
The selected language will be used for all the text shown on the display; and it will be
effective from the next time the monitor is powered up.
NIBP TEST MODE
These menus facilitate performing verification testing for the NIBP subsystem.
For a detailed procedure, refer to the Performance Verification section.
Table 10. NIBP Test
Tests Description
Pressure Sensor
Accuracy Test
Verifies that the pneumatic pressure sensor accuracy is
within the specification.
Air Leakage Test Verifies that the pneumatic pressure air leakage is within
a pressure drop of 6 mmHg/min.
Inflation Speed Test Verifies that the pneumatic pressure inflation is at the
time of 4.0 to 7.5 seconds (to 250 mmHg).
Deflation Speed Test Verifies that the proportional valve will open and bleed off
pressure at the rate of 2.8 to 6.0 mmHg/second
(260mmHg – 30mmHg).
Return Returns to Service Menu.