Mediana P10 Manual de Serviço Página 28

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 27
22
1. Ensure the monitor is set up with the dummy can-large.
2. Ensure NIBP Test screen is active on the monitor, then select “Inflation Time
Measurement” by the jog dial.
3. The monitor displays the pressure of approximately 290 mmHg automatically and
measures the inflation time in seconds.
4. The test result displays at the test completion.
Note: The test will have been successfully completed if the inflation time is 4.0 to 7.5
seconds (to 250 mmHg).
Deflation Rate Measurement
Figure 8. Deflation Rate Measurement
1. Ensure the monitor is set up with the dummy can-small.
2. Ensure NIBP Test screen is active on the monitor, then select “Deflation Rate
Measurement” by the jog dial.
3. The monitor displays the pressure of approximately 290 mmHg automatically, then
measures the deflation rate during reducing the pressure.
4. The test result displays 4 parts (from 260-180mmHg, 180-100mmHg, 100-60mmHg
and 60-30 mmHg) at the test completion.
5. Confirm the result is within the specification.
- 260-180mmHg : 4.8 ~ 6.0 mmHg/s
- 180-100mmHg : 4.8 ~ 6.0 mmHg/s
- 100-60mmHg: 3.5 ~ 5.0 mmHg/s
- 60-30 mmHg : 2.8 ~ 4.2 mmHg/s
Vista de página 27
1 2 ... 23 24 25 26 27 28 29 30 31 32 33 ... 109 110

Comentários a estes Manuais

Sem comentários