9
ROUTINE MAINTENANCE
WARNING: Do not spray or pour any liquid on the monitor or its accessories. Do
not immerse the monitor or its accessories in liquid or clean with caustic or
abrasive cleaners.
Cleaning
The monitor may be surface-cleaned by using a soft cloth dampened with either a
commercial, nonabrasive cleaner or one of the solutions listed below. Lightly wipe the
top, bottom and front surfaces of the monitor.
z 70% Isopropyl alcohol
z 10% Chlorine bleach solution
For cables, sensors, cuffs, transducers and probes, follow cleaning instructions in the
directions for use shipped with those components.
Avoid spilling liquid on the monitor, especially in connector areas. If liquid is accidentally
spilled on the monitor, clean and dry thoroughly before reuse. If in doubt about monitor
safety, refer the unit to qualified service personnel for checking.
Periodic Safety and Functional Checks
The following performance verification tests may be used following repair or during routine
maintenance (if required by your local institution). The following checks should be
performed at least every year by qualified service personnel.
1. Inspect labels for legibility. If the labels are not legible, contact Mediana Technical
Service Department.
2. If the monitor has been visibly damaged or subjected to mechanical shock (for example,
if dropped), perform the performance tests as described in the Performance
Verification section. If the unit fails these performance tests, refer to the
Troubleshooting section.
3. Perform the electrical safety tests described in the Performance Verification section.
If the unit fails these electrical safety tests, do not attempt to repair. Contact Mediana
Technical Service Department.
4. Inspect the fuses for proper value and rating.
Qty 2, 3.15 A, 250 volts for AC mains